ANNEX IV
EC VERIFICATION 1. EC verification is the procedure whereby the manufacturer or his
authorized representative established in the Community ensures and declares that the
products which have been subject to the procedure set out in Section 4 conform to the type
described in the EC type-examination certificate and meet the requirements of this
Directive which apply to them.
2. The manufacturer must take all the measures necessary to ensure that the manufacturing
process produces products which conform to the type described in the EC type-examination
certificate and to the requirements of the Directive which apply to them. Before the start
of manufacture, the manufacturer must prepare documents defining the manufacturing
process, in particular as regards sterilization where necessary, together with all the
routine, pre-established provisions to be implemented to ensure homogeneous production
and, where appropriate, conformity of the products with the type described in the EC
type-examination certificate and with the requirements of this Directive which apply to
them. The manufacturer must affix the CE marking in accordance with Article 17 and draw up
a declaration of conformity.
In addition, for products placed on the market in sterile condition, and only for those
aspects of the manufacturing process designed to secure and maintain sterility, the
manufacturer must apply the provisions of Annex V, Sections 3 and 4.
3. The manufacturer must undertake to institute and keep up to date a systematic procedure
to review experience gained from devices in the post-production phase and to implement
appropriate means to apply any necessary corrective action. This undertaking must include
an obligation for the manufacturer to notify the competent authorities of the following
incidents immediately on learning of them:
(i) any malfunction or deterioration in the characteristics and/or performance of a
device, as well as any inadequacy in the labelling or the instructions for use which might
lead to or might have led to the death of a patient or user or to a serious deterioration
in his state of health;
(ii) any technical or medical reason connected with the characteristics or performance of
a device for the reasons referred to in subparagraph (i) leading to systematic recall of
devices of the same type by the manufacturer.
4. The notified body must carry out the appropriate examinations and tests in order to
verify the conformity of the product with the requirements of the Directive either by
examining and testing every product as specified in Section 5 or by examining and testing
products on a statistical basis as specified in Section 6, as the manufacturer decides.
The aforementioned checks do not apply to those aspects of the manufacturing process
designed to secure sterility.
5. Verification by examination and testing of every product
5.1. Every product is examined individually and the appropriate tests defined in the
relevant standard(s) referred to in Article 5 or equivalent tests must be carried out in
order to verify, where appropriate, the conformity of the products with the EC type
described in the type-examination certificate and with the requirements of the Directive
which apply to them.
5.2. The notified body must affix, or have affixed its identification number to each
approved product and must draw up a written certificate of conformity relating to the
tests carried out.
6. Statistical verification
6.1. The manufacturer must present the manufactured products in the form of homogeneous
batches.
6.2. A random sample is taken from each batch. The products which make up the sample are
examined individually and the appropriate tests defined in the relevant standard(s)
referred to in Article 5 or equivalent tests must be carried out to verify, where
appropriate, the conformity of the products with the type described in the EC
type-examination certificate and with the requirements of the Directive which apply to
them in order to determine whether to accept or reject the batch.
6.3. Statistical control of products will be based on attributes, entailing a sampling
system ensuring a limit quality corresponding to a probability of acceptance of 5 %, with
a non-conformity percentage of between 3 and 7 %. The sampling method will be established
by the harmonized standards referred to in Article 5, taking account of the specific
nature of the product categories in question.
6.4. If the batch is accepted, the notified body affixes or has affixed its identification
number to each product and draws up a written certificate of conformity relating to the
tests carried out. All products in the batch may be put on the market except any in the
sample which failed to conform.
If a batch is rejected, the competent notified body must take appropriate measures to
prevent the batch from being placed on the market. In the event of frequent rejection of
batches, the notified body may suspend the statistical verification.
The manufacturer may, on the responsibility of the notified body, affix the notified
body's identification number during the manufacturing process.
7. Administrative provisions
The manufacturer or his authorized representative must, for a period ending at least five
years after the last product has been manufactured, make available to the national
authorities:
- the declaration of conformity,
- the documentation referred to in Section 2,
- the certificates referred to in Sections 5.2 and 6.4,
- where appropriate, the type-examination certificate referred to in Annex III.
8. Application to devices in Class IIa
In line with Article 11 (2), this Annex may apply to products in Class IIa, subject to the
following exemptions:
8.1. in derogation from Sections 1 and 2, by virtue of the declaration of conformity the
manufacturer ensures and declares that the products in Class IIa are manufactured in
conformity with the technical documentation referred to in Section 3 of Annex VII and meet
the requirements of this Directive which apply to them;
8.2. in derogation from Sections 1, 2, 5 and 6, the verifications conducted by the
notified body are intended to confirm the conformity of the products in Class IIa with the
technical documentation referred to in Section 3 of Annex VII.
|